Examination
Fluoroscopy
10 / 52 AXD3-340.620.13.01.02 Operator Manual
Resetting the fluoroscopic data
Once the examination is ended, then please note the displayed fluoroscopic data
and
◆ Press the reset key.
– The displays for the fluoroscopic data and the area dose product are reset
to zero.
Automatic fluoroscopic control
Selecting automatic fluoroscopic control
The fluoroscopic curves and dose levels preset in the organ program can be
changed with the automatic program keys on the system remote control console.
You can select different fluoroscopic curves with the automatic programs.
◆ You select automatic programs 1, 2 or 3 with one of the keys.
Stop function
If the fluoroscopic values (kV, mA) set by the system should remain constant
◆ Press the stop key during fluoroscopy.
If you want to cancel the stop function
◆ Select one of the 3 automatic programs by a key or an organ program.
Fluoroscopic curves
Depending upon the system configuration, the following fluoroscopic curves are
assigned to the automatic programs:
Automatic
program
Configured with
standard Fluoro
Configured with
standard Fluoro
and SUPERVI-
SION
Configured with
standard Fluoro
and CAREVISION
1 Dose-reduced curve Antiisowatt curve
2 Antiisowatt curve
3 Contrast curve (60 to 80 kV)
1
3
2