Examination
Cassette exposures with wall stand
52 / 52 AXD3-340.620.13.01.02 Operator Manual
Releasing the exposure
❏ To release the exposure, actuate the exposure release button on the genera-
tor ON/OFF console.
❏ Before each exposure release check:
– the collimation with the aid of the light localizer on the multileaf collimator
– the set SID with the tape measure on the multileaf collimator
– that if the dominants are changed the correct dominant is always selected
– the central ray onto the cassette center with the aid of the laser line light
localizer on the multileaf collimator
❏ The signal lamps for radiation ON must light up.
For safety reasons, radiation cannot be released from the system control con-
sole or the footswitch in this mode.
Warning
If these radiation indicators light up although no switch has been actuated:
◆ Switch the system off immediately!
◆ Notify SIEMENS Uptime Service.
◆ Do not start the system up again.