Examination
Cassette exposures with wall stand
52 / 52 AXD3-340.620.13.01.02 Operator Manual
Releasing the exposure
To release the exposure, actuate the exposure release button on the genera-
tor ON/OFF console.
Before each exposure release check:
the collimation with the aid of the light localizer on the multileaf collimator
the set SID with the tape measure on the multileaf collimator
that if the dominants are changed the correct dominant is always selected
the central ray onto the cassette center with the aid of the laser line light
localizer on the multileaf collimator
The signal lamps for radiation ON must light up.
For safety reasons, radiation cannot be released from the system control con-
sole or the footswitch in this mode.
Warning
If these radiation indicators light up although no switch has been actuated:
Switch the system off immediately!
Notify SIEMENS Uptime Service.