POLYDOROS SX 65/80
Integrated Control Console
AXIOM Iconos R200 AXD3-340.620.13.01.02 33 / 36
Selecting an organ program
◆ Actuate the program button in the data menu
– the program menu appears
◆ Select an organ program
– after the selection the system returns to the data menu
– the selected organ program appears inverted
Organ program data indication 0
❏ All data stored for the selected organ program are indicated in the data menu.
Editing organ program data
❏ The exposure data of any organ program can be edited. However, you have to
distinguish between the temporary change and the permanent storage of a
program.
Temporary change 0
◆ Select the organ program
◆ Depending on the activated exposure technique the following exposure data
can be edited:
–kV value
–mAs value
– ms value
– Focal spot
– Tube load
– Measuring fields (dominants)
– Film-screen combination
– Density compensation
In the data menu the selected organ program is displayed non-inverted.
Permanent storage
➩ For further information refer the FLUOROSPOT Compact operating instruc-
tions
Program
Cerv spine
oblique
The edited parameters are in effect only until you select another organ program
or switch the system off.