AXIOM Iconos R200 AXD3-340.620.13.01.02 13 / 54
Accessories
Optional Accessories
Head support
Application Holder for cranial examinations. Can be adjusted in height, width, and lateral tilt
Attachment Move the table to the horizontal position.
Loosen both cap screws (1) for attaching the head support to the accessory
rails.
Insert the head support into the accessory rails at the head end or foot end of
the table and move it to the appropriate position.
Tighten the cap screws (1).
Pull the head support to ensure that it is secured properly.
Turn the following cap screws as indicated to permit adjustment of the:
(2) position from side to side.
(3) height.
(4) lateral tilt from 0° to ± 90° (depending on the height setting).
(5) pads (6) to the width of the patient’s head; turn to the left.
Set the position from side to side.
Tighten the cap screw (2).
Push and pull the head support to ensure that it is seated properly.
Assist the patient in positioning his/her head in the head support.
The patient’s head should make contact with the top pad.
Raise the stabilizing arms to the height required.
Tighten the cap screw (3).
Push and pull on the stabilizing arms to ensure they are seated properly.
Turn the cap screws (5) carefully to the right.
Both padded stabilizing arms will move toward the patient’s head.
2
1
5
1
3
5
1
5
4
1
1
6
6