Examination
Digital Radiography
38 / 52 AXD3-340.620.13.01.02 Operator Manual
Programming run
The X-ray system stands in the starting position (position 1) for the programming
run (abdomen)
Release 1st step
◆ Deflect the joystick above / below or press the key to the left / right
– The X-ray system moves into the next position
– The position is displayed on the display in the remote control console, e.g.
“2/4”
– The settings of the starting position are stored.
◆ Adapt the following parameters with the corresponding operating elements:
– Primary rectangular collimator
– Cu prefilter
–Dominant
– Semitransparent wedge filters (if option present)
Release 2nd step and
further steps
◆ Deflect the joystick above / below or press the key to the left / right
– The X-ray system moves into the next position
– The position is displayed on the display in the remote control console, e.g.
“3/4“ up to max. ”4/4”
– The settings of the positions are stored.
◆ Move to each position one after the other and adapt the radiation field and
dominant as far as required.
Move the X-ray sys-
tem into the starting
position
◆ Deflect the joystick up / down or press the key to the left / right until the X-ray
system has reached the starting position again.
– The position is displayed on the display in the remote control console, e.g.
”1/4”
– The stored parameters are set automatically.
A maximum of 5 steps and thus 6 positions are possible.