AXIOM Iconos R200 AXD3-340.620.13.01.02 17 / 44
Grip locations
The following grip locations are provided:
1 handgrip (front)
Grip protection strip (head end)
Handgrip strip (back)
Ensure that the handgrip strip, the grip protection strip and the handgrip are
always attached.
If these grip locations cannot be used:
Pay special attention to the stated possibilities of crushing between moving
parts and their guide openings.
Ensure during the examination that the patient under no circumstances holds
on to the edges of the patient table.
Not intended as gripping point:
The handle for turning the tube assembly must not be used as grip location or
hold for the patient.
Danger zones
The dotted areas mark the danger zone, where no objects must be located during
tilting.
Prior to activating any movements of the unit, especially tilting of the unit,
make sure that there are no objects such as stairs, steps, stands, waste con-
tainers, instrument tables, beds, gurneys, monitor support systems, operat-
ing consoles or the like in its movement range.
Warning
When handling the system correctly as well as when positioning the patient,
operators and patients should use only the grip locations
provided for this purpose.
Movement range of the positioning table
Danger zone