Examination
Digital Radiography
30 / 52 AXD3-340.620.13.01.02 Operator Manual
Periscanning
Serial exposures of the contrast bolus during aortoarteriography with the patient
at rest.
General
Periscanning is selected after the patient has been positioned, the catheter has
been placed and the system settings for aorto-arteriography have been per-
formed.
Proceeding from the abdomen (starting position), the entire peripheral examina-
tion area is scanned. The scan movement takes place with increasing movement
speed of the X-ray system longitudinal displacement with finely graduated con-
trollable adaptation of the speed to the contrast medium flow. The max. travel
speed of the X-ray system is increased to 16 cm/s.
Selection
Press this key.
Periscanning is selected: the LED lights up.
Select an organ program for Periscanning.
Deselection
Press this key or select another exposure system.
Periscanning is deselected.
You generate an organ program with image series for Periscanning in the Digital
Radiography menu.
DR
DR
The LEDs in the buttons for Periscanning and DR flash if the prerequisites for
Periscanning are not met.
DR
DR