Examination
Free Cassette Exposure
50 / 52 AXD3-340.620.13.01.02 Operator Manual
Releasing an exposure
❏ To release the exposure, actuate the exposure release button on the genera-
tor ON/OFF console.
❏ Before every exposure release check:
– the format limitation with the aid of the light localizer on the collimator
– the set SID with the tape measure on the collimator
– the central ray on cassette center with the aid of the laser line light localizer
on the collimator.
❏ The radiation ON indicators must light up.
For safety reasons, radiation cannot be released from the system control con-
sole or the footswitch in this mode.
Warning
If these radiation indicators light up although no switch has been actuated :
◆ Switch the system off immediately!
◆ Notify the SIEMENS Uptime Service.
◆ Discontinue operation of the system.