Siemens AXIOM Iconos R200 - Reading the Table Tilt

Siemens AXIOM Iconos R200
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Operation
System Settings
20 / 32 AXD3-340.620.13.01.02 Operator Manual
Limitations when tilting in Trendelenburg position:
Tabletop tilting stops when the distance table (Image Intensifier) to floor is less
than 40 mm.
To further tilt the table,
Release the joystick/button and press it once again.
Spotfilm device and tabletop move footwards until table tilting is possible
again;
an acoustic signal sounds during tilting.
Reading the table tilt
System remote
control console
Read the tilt in degrees (“°”) on the display.
Tableside control
panel
Read the display in degrees (“°”).
The movement always stop automatically in the 0° position of the tabletop if the
“Stop in 0° position” function is switched on.
When tilting the table down to a Trendelenburg position >15°, the shoulder sup-
ports supplied or the foot restraints available as optional accessory must be used
to secure the patient. The weight of the patient must not exceed 180 kg.
In the case of a Trendelenburg position >45°, the foot restraints available as op-
tional accessory must be used in addition to the shoulder supports. The weight
of the patient must not exceed 180 kg.
In the case of a Trendelenburg position >45°, the weight of the patient must not
exceed 150 kg.
˚