Examination
Fluoroscopy
8 / 52 AXD3-340.620.13.01.02 Operator Manual
Fluoroscopy with Supervision (optional)
By pressing one of the keys you select
Standard fluoroscopy (Fluoro +) or Supervision (Fluoro -)
Pulsed fluoroscopy (optional)
Press one of the keys briefly to selected
Pulsed fluoroscopy with a pulse rate of 3 (3)
1
p/s, 7.5 (8.3)
1
p/s
or 15 (12.5)
1
p/s or
Standard fluoroscopy.
Increase (Fluoro+) / decrease (Fluoro-) the pulse rate
A pulse rate of 3 p/s is the lowest setting and standard fluoroscopy is the highest
setting.
Roadmap
2
Roadmap can be combined with all fluoroscopic modes. There is a corresponding
Roadmap Fluoro program for every combination.
Press this key to select/deselect Roadmap.
Refer to the FLUOROSPOT Compact Operator Manual for information on the
operating sequence.
Fluoro
-
Fluoro
+
You can also switch from one operating mode to another during a fluoroscopic
examination.
1
Values in brackets apply to VIDEOMED DHC
Fluoro
-
Fluoro
+
You can also switch over from one operating mode to another during fluorosco-
py.
2
only with DSA option
Road-
map
You can select/deselect this function only when radiation is not being released.
Roadmap is automatically deselected when you change to the “Pat. List“ menu,
on changing the exposure system (Peri, bed-side exposure) or the organ pro-
gram selection.