System Operation
On-Off/Emergency Stop
8 / 32 AXD3-340.620.13.01.02 Operator Manual
Switching back on
Remove the hazardous situation.
Unlock the red emergency stop button
The system is immediately ready for use
If the hazardous situation cannot be removed, please notify the Siemens Cus-
tomer Service.
Patient rescue
If the patient has to be rescued in an emergency situation (e.g. power failure),
proceed as follows:
Bring accessories into a position which makes rescue possible.
If the patient is compressed with the cone (optional)
Pull out the knob (1) from the engaged position b and turn it by 90° until it en-
gages in the deep notch (notch position a).
Lift the cone arm off from the patient until it locks in a position of approx. 10°.
a
b
(1)
Warning
If shoulder supports or foot restraints are loosened in Trendelenburg position,
then take care that the operating personnel secure the patient sufficiently.
Obtain the assistance of sufficient personnel for the rescue of the patient.