POLYDOROS SX 65/80
Integrated Control Console
22 / 36 AXD3-340.620.13.01.02 Operator Manual
Select 1-point technique 0
Setting the kV value
Exposure cut-off by automatic exposure control.
Selection Actuate
a button for film-screen combination or
a measuring field (dominant) button
With organ program:
Select an organ program with 1-point technique
Deselection Actuate
the mAs button: 2-point technique is selected or
the ms button: 3-point technique is selected or
the IONTOMAT button: 3-point technique with IONTOMAT is selected
With organ program:
Select an organ program without 1-point technique
Exposure data
to be set
The following exposure data can be set:
–kV value
Focal spot
Tube load
Measuring fields (dominants)
Film-screen combination
Film density compensation
Display of exposure
data
After the end of the exposure, the mAs and ms values in the display are flash-
ing.