POLYDOROS SX 65/80
Integrated Control Console
22 / 36 AXD3-340.620.13.01.02 Operator Manual
Select 1-point technique 0
❏ Setting the kV value
❏ Exposure cut-off by automatic exposure control.
Selection ◆ Actuate
– a button for film-screen combination or
– a measuring field (dominant) button
With organ program:
◆ Select an organ program with 1-point technique
Deselection ◆ Actuate
– the mAs button: 2-point technique is selected or
– the ms button: 3-point technique is selected or
– the IONTOMAT button: 3-point technique with IONTOMAT is selected
With organ program:
◆ Select an organ program without 1-point technique
Exposure data
to be set
❏ The following exposure data can be set:
–kV value
– Focal spot
– Tube load
– Measuring fields (dominants)
– Film-screen combination
– Film density compensation
Display of exposure
data
❏ After the end of the exposure, the mAs and ms values in the display are flash-
ing.
100%
H
U
mAs
ms
kV
40
Program
Iconos
Cassette
0.0
Iconos Bed
ms
Iconos DFR
80%
D
3.4
µGy*cm
2
Info
mA
%hu
kV
45
0.3
0
min0.1
Vertix