POLYDOROS SX 65/80
Integrated Control Console
10 / 36 AXD3-340.620.13.01.02 Operator Manual
Measuring field (dominant) display
If either the left or the right dominant selected becomes invalid, e.g. due to colli-
mation, the measuring field selected is then shown crossed out.
When the diaphragm is opened again, the previously selected dominants are
shown normal again.
Display of the exposure data
After the exposure is completed the actual values of kV, ms and mAs are dis-
played.
Messages
General error messages
Each time the system is switched on, an automatic check of all important func-
tions is made (test routine).
When a fault is detected the generator is blocked and the fault is displayed as er-
ror in the message line and a window with the error message appears on the right
side of the monitor screen.
You can cancel the disable status as follows
– Acknowledge error message.
If the error message is still displayed, then there is an error in the system. Please
call the SIEMENS Uptime Service.
Tube load computer
General
The tube load computer protects the connected X-ray tube assembly from over-
loading in radiographic mode.
From the selected exposure factors and from the physical and geometrical prop-
erties of the tube assembly it computes the length of the cooldown time that may
be necessary.
Furthermore, the tube load computer stores all important operating data of the
respective X-ray tube assembly.
The current load (%hu) is constantly displayed.