Pause time indication
The cooldown time indicated in the message line must be observed to prevent
overloading the tube assembly. The next exposure must not be released before
this time has elapsed.
❏ X-ray tube assembly with no load
– No cooldown time required:
You may release the preset exposure up to 100% tube load.
❏ X-ray tube assembly with little load
– No cooldown time required and automatic switching to 80% tube load.
You may release the preset exposure with 80% tube load.
❏ X-ray tube assembly with heavy load
– Cooldown time is indicated
1
You may release the preset exposure with 80% tube load only after the
indicated cooldown time has elapsed.
Load data indication
The load unit counter stores the load data of the connected X-ray tube assem-
blies.
With the INFO button actuated the load data of the selected X-ray tube assembly
is indicated in the message line.
❏ Tube load units (BE)
– 1 load unit corresponds to 1 exposure or 1s series.
❏ ACQ on wait
– Number of exposures released during the indicated cooldown time.
❏ ACQ on overloads
– Number of overloads.
1
If a warning buzzer is to come on at a certain pause time threshold please have it programmed by
your customer service.