POLYDOROS SX 65/80
Integrated Control Console
AXIOM Iconos R200 AXD3-340.620.13.01.02 11 / 36
Pause time indication
The cooldown time indicated in the message line must be observed to prevent
overloading the tube assembly. The next exposure must not be released before
this time has elapsed.
X-ray tube assembly with no load
No cooldown time required:
You may release the preset exposure up to 100% tube load.
X-ray tube assembly with little load
No cooldown time required and automatic switching to 80% tube load.
You may release the preset exposure with 80% tube load.
X-ray tube assembly with heavy load
Cooldown time is indicated
1
You may release the preset exposure with 80% tube load only after the
indicated cooldown time has elapsed.
Load data indication
The load unit counter stores the load data of the connected X-ray tube assem-
blies.
With the INFO button actuated the load data of the selected X-ray tube assembly
is indicated in the message line.
Tube load units (BE)
1 load unit corresponds to 1 exposure or 1s series.
ACQ on wait
Number of exposures released during the indicated cooldown time.
ACQ on overloads
Number of overloads.
1
If a warning buzzer is to come on at a certain pause time threshold please have it programmed by
your customer service.
100%
H U
mAs
ms
kV
66
Program
Iconos
Cassette
0.0
Iconos Bed
ms
Iconos DFR
80%
D
mAs
0.20
ms3.3
3.4
µGy*cm
2
Info
mA
%hu
kV
45
0.3
0
min0.1
Vertix
1234567
0
0
Tube load units:
Tube No.: 1 ACQ on wait
ACQ on overloads
INFO

Table of Contents

Related product manuals