POLYDOROS SX 65/80
Integrated Control Console
12 / 36 AXD3-340.620.13.01.02 Operator Manual
1%hu Indication of the current value of the tube load in relation to the maximum heat
storage capacity in per cent.
Functional and safety checks
Daily checks
After switching on
◆ Please make a visual inspection of the displays and signals on the integrated
generator control console.
No error message may appear.
The radiation warning lights in the room must not light up.
During the examination
◆ Check the radiation indicator. It may light up only during the duration of the X-
ray exposure.
Monthly checks
◆ Perform a functional check of the automatic exposure control, as described in
Register “Safety”.
Legally required checks
Obtain information in Register “Safety”.
Maintenance intervals
Maintenance must be performed annually to maintain the safety and functioning
of the generator.
If you have not concluded a maintenance contract, please notify the SIEMENS
Uptime Service in good time.
INFO button is not active if an information is already present on the message
line.
If a radiation warning light lights up without actuating a switch
◆ immediately switch off the X-ray system through emergency off and
◆ notify the SIEMENS Uptime Service.