POLYDOROS SX 65/80
Integrated Control Console
AXIOM Iconos R200 AXD3-340.620.13.01.02 25 / 36
Select 3-point technique with IONTOMAT 0
❏ Setting the kV value and the maximum mAs and ms values
❏ Exposure cut-off by the automatic exposure control.
Selection ◆ Actuate the IONTOMAT button
With organ program:
◆ Select an organ program for 3-point technique with IONTOMAT
Deselection ◆ Actuate
– the IONTOMAT button: 1-point technique is selected
With organ program:
◆ Select an organ program without 3-point technique with IONTOMAT
Exposure data
to be set
❏ The following exposure data can be set:
–kV value
–mAs value
– ms value
– Focal spot
– Tube load
– Measuring fields (dominants)
– Film-screen combination
– Film density compensation
Display of exposure
data
❏ After the end of the exposure, the mAs and ms values in the display are flash-
ing.