AXIOM Iconos R200 AXD3-340.620.13.01.02 49 / 52
Examination
Free Cassette Exposure
Collimation on exposure
The automatic format collimation of the system is not effective if you have se-
lected ”tabletop cassette exposure or free mode”, or if you have turned the
tube assembly into a position outside 0°.
You can set the exposure format freely.
The limits of the maximum radiation field are determined by the mecha-
nism of the primary collimator.
The collimator does not react to a change of the SID.
The digital display field of the primary collimator shows ”Manual”.
Selecting the exposure technique
Tabletop cassette exposure or free mode
Press this key
The LED lights up
With tabletop cassette exposure or the tube assembly swivelled out, the sys-
tem automatically switches over to “Free mode / tabletop exposure”. The au-
tomatic format collimation system is disabled.