AXIOM Iconos R200 AXD3-340.620.13.01.02 5 / 32
System Operation
On-Off/Emergency Stop
Switching the system on
◆ Press the button on the generator ON/OFF console
1
.
– The entire system and all further connected devices are switched on.
– A safety test program runs automatically from this time.
– The welcome menu appears on the monitor.
The system is ready after initialization and the end of the safety test program if
at the integrated
generator control
console
❏ no error message appears in the message line of the data menu,
❏ none of the kV, mAs or ms displays flashes,
❏ a system selection is active,
on the live monitor ❏ the patient menu appears,
on the system remote
control console
❏ no displays flash,
❏ no error message appears,
on the tableside con-
trol panel
❏ no displays flash,
Emergency stop ❏ no red emergency stop button is activated.
In the case of a fault of the safety test program switch the system off and back
on within 10s. The system again starts with the automatic safety test program.
As a rule the fault is rectified by this.
If this is not the case, switch the system off. After the image system has auto-
matically switched off (monitor dark), wait 30 s and then switch it on again. In this
case the image system is run down and restarted. A possible error will be dis-
played on the monitor.
Please notify the SIEMENS Uptime Service if this is the case.
➩ Refer to the FLUOROSPOT Compact Operator Manual for further information.
Perform a functional and safety check according to the Functional and Safety
Check chapter.
1
If a local hardcopy camera connection is present, the associated camera and network components
should be switched on first.