AXIOM Iconos R200 AXD3-340.620.13.01.02 5 / 32
System Operation
On-Off/Emergency Stop
Switching the system on
Press the button on the generator ON/OFF console
1
.
The entire system and all further connected devices are switched on.
A safety test program runs automatically from this time.
The welcome menu appears on the monitor.
The system is ready after initialization and the end of the safety test program if
at the integrated
generator control
console
no error message appears in the message line of the data menu,
none of the kV, mAs or ms displays flashes,
a system selection is active,
on the live monitor the patient menu appears,
on the system remote
control console
no displays flash,
no error message appears,
on the tableside con-
trol panel
no displays flash,
Emergency stop no red emergency stop button is activated.
In the case of a fault of the safety test program switch the system off and back
on within 10s. The system again starts with the automatic safety test program.
As a rule the fault is rectified by this.
If this is not the case, switch the system off. After the image system has auto-
matically switched off (monitor dark), wait 30 s and then switch it on again. In this
case the image system is run down and restarted. A possible error will be dis-
played on the monitor.
Please notify the SIEMENS Uptime Service if this is the case.
Refer to the FLUOROSPOT Compact Operator Manual for further information.
Perform a functional and safety check according to the Functional and Safety
Check chapter.
1
If a local hardcopy camera connection is present, the associated camera and network components
should be switched on first.