System Operation
System Settings
AXIOM Iconos R200 AXD3-340.620.13.01.02 23 / 32
Switching on the fulcrum height light localizer
◆ Press one of the keys briefly.
– The light localizer goes out automatically after approx. 20s.
Reading the fulcrum height
The fulcrum height display on the system remote control console is decisive for
all examinations.
System remote
control console
◆ Read the display (display in ”mm”).
on the system Switch the fulcrum height light localizer on.
You have 2 possibilities of reading the fulcrum height:
Direct method (light line on the object):
◆ Look from the right side in the mirror on the system.
– You will see the projection of a red laser light line directly on the patient.
– This light line indicates the set fulcrum height directly on the patient.
Indirect method (on scale):
◆ Look at the scale next to the mirror on the system.
– You will see a red light mark there.
◆ Read the scale value indicated by the light mark.
– One division corresponds to 0.5 cm.
Press briefly and the fulcrum height changes by 1 mm.
Press longer, the fulcrum height changes continuously in steps of 1 mm.