Examination
Digital Radiography
44 / 52 AXD3-340.620.13.01.02 Operator Manual
Examination procedure for images of the spine and ortho-
pedic longleg images
Preparation
Prepare the digital imaging system.
Select the DR system at the unit remote control console.
Position the patient table vertically (+ 90°).
Have the patient stand in the unit (with/without footrest) and immobilize the
patient with the compression belt.
For reasons relating to the exposure technology, the scanning direction for DR
scanning is
from bottom to top (towards the head end) for spinal acquisitions
from top to bottom (towards the foot end) for orthopedic whole leg acquisi-
tions
Set horizontal projection.
Set the largest possible SID.
Press this key.
The LED in the key lights up if the conditions for scanning are not fulfilled.
DR scanning is selected.
I.I. full format is selected.
The rectangular height diaphragm moves automatically to the configured
value (approx.10 - 12 cm)
The display shows 3 scanning programs with 2 directions each.
The selected scanning program is highlighted.
By pressing the key again the next possible scanning program with direction be-
comes active.
Select a scanning program for spine or orthopedic longleg images
left arrow: scanning direction from bottom to top
right arrow: scanning direction from top to bottom
If the LED in the start key flashes,
press the key.
The compression device moves into park position, if present.
Oblique setting moves into the 0° position.
Once these positions are reached, the flashing LED goes out.
DR
Ln
S
i
pne<- -
gelg-->o<- -
<- - Abdomen - ->
-->
Start