Accessories
Optional Accessories
34 / 54 AXD3-340.620.13.01.02 Operator Manual
Left: Releasing fluoroscopy Right: Releasing radiography
Releasing fluoroscopy
◆ Step on the right pedal (foot switch for fluoroscopy). Continue pressing down
on the pedal as long as fluoroscopy should remain active.
◆ Let go of the pedal.
❏ Fluoroscopy then ends.
Releasing radiography
◆ Step on the left pedal (foot switch for radiography). Continue to press down
on the pedal until the exposure has ended.
❏ The exposure has ended when the radiation ON indicator on the remote con-
trol console goes out.
◆ Let go of the pedal.
Removal
◆ Turn the locking ring on the cable plug to the left.
◆ Remove the plug from the receptacle.
◆ Roll up the foot switch cable.
◆ Store the foot switch and cable in an appropriate location.
Depending on the frequency of use, you can leave the foot switch attached or
disconnect it after the examination.