Accessories
Optional Accessories
AXIOM Iconos R200 AXD3-340.620.13.01.02 37 / 54
Insert the radiation shield in the front accessory rail and slide it to the appro-
priate work position.
Tighten both cap screws.
Push and pull on the radiation shield to ensure that it is seated properly.
Tilt the radiation shield into a protective position.
The radiation shield can be tilted inward on the accessory rail as well as out-
ward and all the way down.
Removal Move the patient table to the horizontal position.
Tilt the radiation shield to the vertical position.
Loosen both cap screws.
Remove the lateral radiation shield from the front accessory rail.