Accessories
Optional Accessories
48 / 54 AXD3-340.620.13.01.02 Operator Manual
◆ Be sure to set the examination unit horizontal (± 10°)!
◆ As required hook a lead-rubber apron on the front bar of the holder.
◆ Hook the second lead-rubber apron according to the position of the examiner
on the left or right bar.
Removal ◆ Unhook the lead-rubber aprons from the bars after one another and carefully
place them on the side (do not kink them).
◆ Pull the bars out from the accessory rails of the collimator.
◆ Store the accessory in a suitable place.
Radiation protection
value:
The lead equivalent of the double-layer lead-rubber apron is 0.5 mm Pb at 80 kV.
Do not exceed the maximum permissible additional weight of 70N or 7kg. This
means that at the most 2 lead-rubber aprons may be hooked onto the bar.
The lead-rubber elements of this accessory are inserted in welded plastic pock-
ets. These pockets are riveted with the rails.
The plastic pockets can be cleaned easily with mild dish-washing liquid solu-