Technical Description
Technical Data
12 / 12 AXD3-340.620.13.01.02 Operator Manual
Fluoroscopy Working range:
Normal fluoroscopy: from 40kV/0.2mA to 110kV/4.1mA
High-contrast fluoroscopy: from 40kV/0.2mA to 110kV/18mA
maximum of 4 FLUO curves can be configured
Tube voltage 53 values from 40 to 150 kV
In steps of half Siemens exposure points (EP) or 1 EP (configurable)
Automatic system 0-,1- point technique with continuously falling load
2-, 3- point technique with constant load
3- point technique (with IONTOMAT PL) with constant load
mAs integrator 65 values from 0.5 mAs to 800 mAs
In steps of half Siemens exposure points (EP) or 1 EP (configurable)
Switching time 1- point technique: 1 ms to 5 s with mAs and time postindication
2- point technique: 2 ms to 5 s depending upon kW, mAs and kV
3- point technique: 20 ms to 5 s depending upon kW, mAs and kV in 49 steps
In steps of half Siemens exposure points
Tube assembly
connection
2 Siemens double-focus tube assemblies
Number of
workstations
max. 5
Automatic exposure
control
IONTOMAT PN with 6 measuring inputs for measuring chambers as well as mul-
tiplier connection (option) or photodiode array (PDA)
Tolerances kV accuracy +/- 5%
mAs accuracy +/- 5% or +/- 0.5 mAs, depending on which value is higher.
mA accuracy +/- 5% or +/- 0.1 mA in fluoroscopy, depending on which value is
higher
mA accuracy +/- 10% in radiography