Ionics Instruments 2004 Page 6-15 DLM 30007-08 Rev. A
To turn the main power off when running on-line:
1. Press ENTER to display the RunTime menu.
2. Select STOP TOC.
3. Press ENTER. Once the MAIN menu is displayed, turn the power off.
To turn the main power off when running grab samples:
1. Use the arrows ( or ) to scroll to STOP in the Grab Sample menu.
2. Press ENTER.
3. When the Grab Sampling Idle menu is displayed, turn off the power.
To turn the main power off when running the autosampler:
1. Select Stop from the File menu of the DataPro Autosampler Program.
2. When the hourglass disappears, turn the power off.
Grab Sample Measurements
The analyzer can also be used for measurement of TOC, TC, and IC directly
from sample bottles. In the Grab Sample mode, specify how many replicate
measurements to make from a sample bottle and how many of these
measurements are to be used in the calculation of average values and
standard deviation or coefficient of variation (relative standard deviation).
WARNING
Make sure the DI Reservoir is full, particularly when running samples with high
TOC’s or high salt concentrations. If there is an air bubble in the reservoir,
high levels of salts may contaminate the reservoir and deplete the resin. Always
“clean-up” the analyzer by running low TOC DI water (at least three
measurements) after running high TOC or salt samples.
If using the analyzer for on-line measurements, change the sample mode in
the SETUP menu to GRAB SAMPLING and change the sample inlet plumbing.
If running both on-line and grab samples, consider using the sample switching