Ionics Instruments 2004 C-5 DLM 30007-08 Rev. A
Setting Alarms
Alarms may be activated to indicate that TOC measurement has been
interrupted when syringes are being filled.
To activate this alarm setting:
1. From the SETUP menu, select OUTPUTS.
2. Select ALARMS.
3. Select RUN ALARM .
4. Select ON.
The alarm display will read TOC Active. An alarm will respond when the TOC
measurements are interrupted.
Calibration in Turbo Sampling Mode
Calibration of Turbo mode is similar to calibration in the normal operating
modes. Before attempting to calibrate the analyzer in TURBO SAMPLING
MODE, review the Calibration chapter. Differences in the specific calibration
procedures are provided below.
NOTE
The calibration constants for Turbo mode are separate from the constants for
other modes and should not be interchanged.