Ionics Instruments 2004 Page 4-18 DLM 30007-08 Rev. A
OXIDIZER
A
CID
LAMP
PUMP TUBING
ENTER
CLEAR
REAGENT FLUSH
PRINT REPORT
FIGURE 4-13: CONSUMABLES Menu
WARNING
The procedures described below reset timers and other key indicators used to
keep track of the Acid, Oxidizer, UV Lamp, and Pump Tubing lifetimes. In
these options, pressing ENTER zeroes the timers. Do not reset the timers unless
the consumables are replaced. When viewing the status of the Acid, Oxidizer,
UV Lamp, and Pump Tubing, press CLEAR to exit the menu without resetting
the timers.
Oxidizer Reagent
To enter the installation date for the oxidizer reagent:
1. From the CONSUMABLES menu, use the arrows ( or ) to scroll
to OXIDIZER. Press ENTER to display the OXIDIZER
INSTALLATION menu.