Ionics Instruments 2004 Page 10-4 DLM 30007-08 Rev. A
Corrective Actions
Actions to be taken if warnings and errors occur are outlined below. If you
need further assistance, contact Ionics Instruments at 1-800-255-6964 or 303-
444-2009.
Warnings 1-4: The analyzer is operating outside its allowable range of
10 °C to 40 °C. Change the ambient temperature or move the analyzer.
Errors 5-6: There is a bad thermistor or electronics problem. Contact Ionics
Instruments for repairs.
Warnings 7-8: Either there is a gas bubble in the conductivity cells, the DI
resin bed is exhausted and needs to be replaced, or there is an electronics
problem. In some cases, refilling the DI reservoir and running the analyzer on-
line or off-line, with the cap removed from the DI water reservoir will remove
the bubble and eliminate the error message. Otherwise, contact Ionics
Instruments.
Errors 9-10: There are problems with the conductivity electronics or cell or a
gas bubble in the conductivity cells (see Warning 7).
Warning 11: Divide by zero error or other calculation error due to electronics
problem, contact Ionics Instruments.
Warnings 12-16: Reagent supplies, lamp, or tubing needs to be replaced.
Follow the procedures described in the Maintenance Chapter, for detailed
instructions.
Warnings 17-25: There is a problem with the electronics or a gas bubble in
the conductivity cells. In some cases, refilling the DI reservoir and running the
analyzer on-line or off-line, with the cap removed from the DI water reservoir
will remove the bubble and eliminate the error message. Otherwise, contact
Ionics Instruments.