Ionics Instruments ļ›™2004 Page 4-11 DLM 30007-08 Rev. A
Printer Cable
The analyzer has a 25-pin parallel (Centronics) printer port for the direct output
of results to a printer. Connect the proper cable for the printer, optional, to the
printer port (see Figures 4-4 or 4-5) and secure with screws. The cable is a
standard computer-printer cable and can be found at local computer stores.
The printer must be IBM compatible or Epson FX 850 compatible. See
Appendix B for CE Mark Specifications.
Installing Reagent Cartridges
The oxidizer and acid reagents have been shipped from Ionics Instruments in
specific packaging. Carefully read the attached MSDS sheets prior to opening
packaging. Within the packaging are two reagent cartridges. Each reagent
cartridge is identified with bright yellow labels that indicate which cartridge
contains Ammonium Persulfate and which cartridge contains Phosphoric Acid.
Warning
Installation of the reagents requires access to the inside of the analyzer. To
avoid potentially dangerous shock, disconnect the power cord before opening
the analyzer's top panel.
Warning
To avoid exposure to the chemical reagents, wear acid-resistant gloves,
protective clothing, and safety goggles or a face shield when changing the
reagent supplies.
WARNING
Reagent containers are for single-use only – do not refill. Refilling or reusing
reagent containers will void all instrument and parts warranties and nullify any
performance claims.
1. Turn off the main power switch and unplug the power cord.
2. Open the analyzer top panel by pushing back on the two black
handles and lifting the panel upwards.