EasyManua.ls Logo

Sievers TOC 800 - Manual;Grab Flow Rates

Default Icon
255 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ionics Instruments 2004 Page 4-26 DLM 30007-08 Rev. A
If one of the preset flow rate options is selected, the name of the selected
option will be displayed in the lower right-hand side of the TOC Measurement
Display.
Manual/Grab Flow Rates
The preset flow rates are only for the on-line sampling mode. If one of the
preset flow rates for on-line sampling is not selected, or if using the grab
sampling mode, the acid and oxidizer flow rates must be set according to
Table 2 as described below. If the preset flow rates for on-line sampling have
been selected, proceed to the printer setup section.
To change the flow rates for on-line or grab sampling from the SETUP menu:
1. Press the down arrow () to scroll to REAGENTS. Press ENTER.
The REAGENTS menu is displayed:
ENTER
CLEAR
ONLINE FLOW RATES
GRAB FLOW RATES
REAGENT FLUSH
FIGURE 4-19: REAGENTS Menu
2. Use the arrows ( or ) to select on-line flow rates or grab flow
rates, as applicable and press ENTER. The flow rates menu for that
sampling mode will be displayed.

Table of Contents

Related product manuals