Ionics Instruments 2004 C-3 DLM 30007-08 Rev. A
FIGURE C-3: SAMPLE MODE Menu – Turbo Analyzer (TURBO SAMPLING selected)
NOTE
Turbo sampling is only available in on-line mode.
Setting Reagent Flow Rates
NOTE
The reagent flow rates for Turbo mode are preset and should be only as follows:
Acid = 2.0 µl/min
Oxid = 0.5 µl/min
These flow rates are optimized for operation from 10-2500 ppb TOC.
The REAGENTS menu for the Sievers 800 Series Turbo analyzer contains an
option to select reagent flow rates for fast sampling.
To set TURBO flow rates:
1. Select SETUP from the MAIN menu.
2. Use the arrows ( or ) to scroll to REAGENTS and press ENTER.
ONLINE SAMPLING
GRAB SAMPLING
ENT = SELECT, CLR = EXIT
TURBO SAMPLING
CURRENT MODE: TURBO
CLEAR
ENTER