EasyManuals Logo

Sievers TOC 800 User Manual

Default Icon
255 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #227 background imageLoading...
Page #227 background image
Ionics Instruments 2004 C-3 DLM 30007-08 Rev. A
FIGURE C-3: SAMPLE MODE Menu – Turbo Analyzer (TURBO SAMPLING selected)
NOTE
Turbo sampling is only available in on-line mode.
Setting Reagent Flow Rates
NOTE
The reagent flow rates for Turbo mode are preset and should be only as follows:
Acid = 2.0 µl/min
Oxid = 0.5 µl/min
These flow rates are optimized for operation from 10-2500 ppb TOC.
The REAGENTS menu for the Sievers 800 Series Turbo analyzer contains an
option to select reagent flow rates for fast sampling.
To set TURBO flow rates:
1. Select SETUP from the MAIN menu.
2. Use the arrows ( or ) to scroll to REAGENTS and press ENTER.
ONLINE SAMPLING
GRAB SAMPLING
ENT = SELECT, CLR = EXIT
TURBO SAMPLING
CURRENT MODE: TURBO
CLEAR
ENTER

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sievers TOC 800 and is the answer not in the manual?

Sievers TOC 800 Specifications

General IconGeneral
BrandSievers
ModelTOC 800
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals