Ionics Instruments 2004 Page 10-5 DLM 30007-08 Rev. A
Warning 26: The write-protect tab on the top of the RAM card is in the protect
position. Disconnect the AC power cord, open the top of the analyzer, and
move the tab to the unprotected position (see Figure 11-2, the tab should be
positioned away, towards the back of the analyzer).
Warning 27: The RAM card is full of TOC data which will be overwritten and
data will be lost if the card is left in the analyzer. To archive data on the RAM
card or have not downloaded the information, turn the analyzer off, remove the
RAM card, and replace with a new card. To archive the data on the RAM
card, but still download the data, turn the analyzer off, remove the card,
download the data using the RAM card reader (see the RAM card chapter),
and reinstall the card.
Warning 28: There is a checksum error. To avoid losing the data stored on
the RAM card, disconnect the AC power cord, remove the card from the
analyzer and download the data with the RAM card reader. Replace the RAM
card, and if the error continues, contact Ionics Instruments.
Warnings 29-30: The RAM card battery is running low or is dead. The data
stored on the card may be lost, if the card is removed from the analyzer, and
wait before downloading the data. While the card is in the analyzer, use the
RS-232 to download the data. the data may also be downloaded via a RAM
card reader if the card is read immediately. the data has been collected, install
a new battery in the card. The procedure for replacing the battery is in the
Ram Card Chapter.
Warning 31: The RAM card is not in the analyzer and data will not be stored.
Install a RAM card to store the data. The procedure for installing the card is in
the Ram Card Chapter.
Errors 32-34: There is a problem with the servomotors used to refill the
syringes or an electronics problem. Stop the analyzer, remove the AC power
cord, and open the top of the analyzer. Check if there is an obstruction in the
mechanism and, if there is, remove the obstruction. Check to make sure that
the ribbon cable located to the left of the syringe pump PC board is plugged
into the sockets. Check the two black connectors on the wiring to the syringe