Ionics Instruments 2004 Page 9-8 DLM 30007-08 Rev. A
this range, proceed to Step 5. If it does not fall within this range, repeat
Steps 3 and 4.
Step 5 – Turn on UV lamp:
After the calibration values have been entered, the UV lamp must be turned
on before using the instrument.
To turn on the lamp:
1. Press CLEAR to return to the MAIN menu, scroll to SETUP and press
ENTER.
2. Scroll to UV LAMP and press ENTER.
4. Use the arrows
( or ) to change the state to ON, press ENTER to set
the state and press CLEAR twice to return to the MAIN menu.
TC Calibration Constant
Use Ionics Instruments-prepared standards or prepare calibration standards
on site according to the procedure in Appendix E.
Step 1 – Configure analyzer for standard measurements and measure reagent
water blank:
1. Perform a REAGENT FLUSH as described in the Installation chapter.
Select BOTH to flush both the ACID and the OXIDIZER reagent
cartridges.
2. From the Main menu, scroll to SETUP and press ENTER.
3. Use the
arrows ( or ) to scroll to OUTPUTS and press ENTER.
4. From the OUTPUTS menu, use the
arrows ( or ) to scroll to
DISPLAY and press ENTER.