Ionics Instruments ļ2004 Page 9-20 DLM 30007-08 Rev. A
Follow the TC Calibration Constant procedure and record the values on the
Verification Worksheet.
Step 2 ā Measure the IC Standard:
Follow the IC Calibration Constant procedure and record the values on the
Verification Worksheet.
This completes the calibration of the instrument. Use the following form to
record the values of the new TC and IC Calibration Constants and the new TC
Zero Offset. Print the list of constants from the SETUP/OUTPUTS/PRINTER
menu. Reset the display to show TOC values under the
SETUP/OUTPUTS/DISPLAY menu.