Ionics Instruments 2004 Page 5-4 DLM 30007-08 Rev. A
For example, the data stream for an on-line measurement will have the form:
ONLINE,,,,2/15/95,13:21,24400.00,218.00,24618.00,0.50,0.30,,,
Sample Mode
Not Used
Date
Time
TOC (ppb)
IC (ppb)
TC (ppb)
Oxidizer Flow Rate (µL/min)
Acid Flow Rate (µL/min)
Error
Messages
1.25,1.25,0.00,60 HZ,UV ON
TC Transfer Constant
IC Transfer Constant
TC/IC Offset
AC Frequency
UV Lamp State
The data stream for a grab sampling measurement will have the form:
GRAB,,2,3,2/15/95,13:21,24400.00,218.00,24618.00,0.50,0.30,,,
Sample Mode
Not Used
Date
Time
TOC (ppb)
IC (ppb)
TC (ppb)
Oxidizer Flow Rate (µL/min)
A
cid Flow Rate (µL/min)
Error
Messages
Sample #
Repetition
1.25,1.25,0.00,60 HZ,UV ON
TC Transfer Constant
IC Transfer Constant
TC/IC Offset
AC Frequency
UV Lamp State