Ionics Instruments 2004 Page 5-6 DLM 30007-08 Rev. A
If a PC is connected to the analyzer, start and stop TOC measurements and
download the RAM card data using a communication program and a two-
letter code.
Set the serial port to: No Parity, 8 data bits, 1 stop bit.
Action Desired Command
To start the analyzer in the on-line mode type TB and press ENTER
To stop TOC measurement: type TE and press ENTER
To download data from the RAM card* type GD and press
ENTER
• Refer to Downloading the RAM Card Data for more information.
To start the TOC Analyzer in the ONLINE mode from the PC, type TB (or tb)
and press ENTER. The analyzer will respond with a TB echo if the command
was successfully executed. The TB command (TOC Begin will start the TOC
measurement cycle using the settings for reagent flow rates output, etc. as
they were when the analyzer was last stopped.
NOTE
The TB command can only be used in the on-line mode. The analyzer will
respond with ER (error) if the system is in the Grab, Multistream, or
Autosampler mode.
To stop the TOC Analyzer by typing TE and pressing ENTER. It will respond
with a TE echo if the command was successfully executed. The TE
command (TOC End) will stop the analyzer and change the display to the
MAIN menu.