CHAPTER 7 Quality control
7-4 Sysmex XE-5000 Instructions for Use
Revised April 2007
4. Perform X-barM setting.
The setting parameters are as follows:
X-barM: Sets whether or not to use X
M control.
Click the “Control” radio button to use
and the “Not Control” radio button not to
use this QC.
X-barM Setting:
Sets the number of samples in a batch by
discrete.
5. After completing settings, click on “OK”, “Cancel” or
“Apply”.
OK: Sets the new settings and closes the
dialog box.
Cancel: Cancels the new settings and closes the
dialog box.
Apply: Sets the new settings.
Limit setting
1. Start quality control from the Information Processing Unit
(IPU).
2. Click the “Setting” button.
The QC Setting dialog box will be displayed.
3. Click the “Limit Setting” tab.
The Limit Setting screen will be displayed.