CHAPTER 9 Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use 9-15
Revised April 2007
4. Loosen the fixing screw at bottom of the detector
chamber, remove the sheath nozzle.
5. Apply CELLCLEAN on the brush provided, and dab by
lightly prodding the brush against the aperture.
6. Insert the sheath nozzle into the detector chamber, and
tighten the fixing screw.
Warning!
Never touch the detector when the power of the Main
Unit is turned ON.
Otherwise, electrical shock may result.
Caution!
• When removing the sheath nozzle, reagent can leak
from the detector chamber. Place gauze or the like
under the detector chamber.
Otherwise, leakage or electrical shock may result.
• As a sheath nozzle is easy to bend, do not drop or
apply impact on it. When removing the detector
chamber or sheath nozzle, take care not to apply
excessive force to the tube that is connected to the
detector chamber.
Otherwise, analysis may not be performed correctly.
Caution!
Be sure to use CELLCLEAN only.
Note:
After using the brush, wash it in water thoroughly to
remove CELLCLEAN before storing it.
Caution!
When closing the detector cover, take care not to bend
the tube.
Otherwise, analysis may not be performed correctly.
Sheath
nozzle
Fixing screw
Loosen