CHAPTER 9 Cleaning and Maintenance
9-22 Sysmex XE-5000 Instructions for Use
Revised April 2007
Entry ranges for reagent amounts
5. After entry is completed, select “OK” or “Cancel” on the
function menu.
OK: The entered contents are saved, and the
screen returns to the Reagent setting screen.
When the Reagent setting screen reappears,
“Replace” automatically appears at the
entered reagent to designate it as a
replacement.
Cancel: The entered contents are deleted and the
screen returns to the Reagent setting screen.
Note:
If a value is entered that exceeds the entry range, the
value will be corrected to a value within the entry
range when “OK” is selected from the function menu
in step 5.
Reagent
name
Entry range
Reagent
name
Entry range
EPK
0
99
FFS
0999
ESE
0.0
99.9
SIM
0.099.9
SLS
0.099.9
SNR
0.099.9
FBA
0.099.9
RED
0.099.9
FFD
0.0
99.9
Information
When “OK” is selected, a check is made using the
dates entered for the expiry date and expiry date after
reagent is opened. If any reagents have passed the
entered expiry limit, the Expired Reagent screen
appears.
Select “OK” on the function menu of the Expired
Reagent screen to return to the previous screen.
Note:
Selecting “Replace” on the function menu alternately
designates the reagent as a replacement and
removes the designation.
Next No.1
DP No.
Num
DP
Xm
C D N R
<EPK>
Manual
Not Ready Replace EPK
OK
This reagent has expired
Exp. Date:2001/11/11
<EPK>