CHAPTER 9 Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use 9-21
Revised April 2007
Registering reagent information (manual entry)
1. Display the Reagent setting screen.
2. Press the and keys, or the number keys, to select a
reagent for manual entry.
3. Select “Manual on the function menu.
The Manual Input screen appears.
4. Press the and keys to select the lot No., expiry date,
expiry date after reagent is opened or container amount,
then enter the appropriate information.
Lot No.: A maximum of 8 alphanumeric characters
can be entered.
This column may also be left blank.
Exp. Date: Only numbers can be entered for this item.
Press the and keys to change
among the year, month, and day areas.
This column may also be left blank.
Exp. After Opened:
A number from 0–99 can be entered for
this item.
This column may also be left blank.
Amounts: Numbers and a decimal point (period) can
be entered for this item.
This column may also be left blank.
Note:
While the Manual Input screen is open, it is also
possible to use the handheld barcode scanner to read
the barcodes that are affixed to the reagent.
A barcode can also be read for a reagent that is
different from the screen that is currently displayed.
When a different reagent’s barcode is read, the screen
changes to the screen for that reagent, and the
barcode information appears. If the previous screen
contents had been changed, the changes are
automatically saved before the screen changes.
Information
After an expiry date, or expiry date after reagent is
opened is entered, and the or key is pressed, a
date check is performed before the cursor moves. If
the entered date is invalid, a beep sounds and the
cursor does not move.
Next No.1
DP No.
Num
DP
Xm
C D N R
OK Cancel
<Manual Input>
<EPK>
ZA0005
Manual
Not Ready Replace EPK
Reagent
Lot No
Exp. Date 2001/11/11
Exp.After Opened 30 Days