CHAPTER 9 Cleaning and Maintenance
9-28 Sysmex XE-5000 Instructions for Use
Revised April 2007 EU
7. Remove the cap of a new RET SEARCH (II) dye solution,
insert the pipe straight into it, and tighten the cap.
8. Insert it all the way into the holder.
9. Close the Main Unit front cover.
10. Press the HELP key on the Main Unit panel keypad. The
HELP screen will appear.
11. If you select “OK” on the function menu of the HELP
screen, the Reagent setting screen will appear.
See “Chapter 9: 9.13 Replacing and registering reagents”
for explanation of the Reagent setting screen.
Information
Insert the RET SEARCH (II) dye solution all the way
into the holder.
If it slants, air bubbles could enter, making it
impossible to obtain correct analysis results.