Revised April 2007 EU
2. Hydraulic system block diagram
WBC/RET
RBC/HGB
0.9 mL
18 µL
18 µL
18 µL
0.9 mL
0.9 mL
0.9 mL
0.6 mL
18 µL
18 µL
4.5 µL
18 µL
2.4 µL
M
CELLPACK Optical detector block
STROMATOLYSER-4DS
STROMATOLYSER-NR (S)
dye solution
RET SEARCH (II)
dye solution
STROMATOLYSER-4DL
Reaction chamber (1:51)
STROMATOLYSER-NR (L)
lyse reagent
Reaction chamber (1:51)
RET SEARCH (II) diluent
Reaction chamber (1:204)
STROMATOLYSER-FB
Reaction chamber (1:50)
STROMATOLYSER-IM
IMI detector chamber (1:250)
SRV
Sampler 200 µL
Manual 130 µL
Whole blood
aspiration pump
FCM
sheath injector
piston
M
2.1 mL
2 mL
1 mL
0.5 mL
3.0 µL
4.0 µL
SULFOLYSER
HGB Flow Cel (1:500)
SRV
CELLPACK
RBC detector 11.7 µL analysis
Hydro Dynamic Focusing (DC Detection)
RBC sample chamber (1:500)
CELLSHEATH
Sheath injector
piston
Wavelength: 555 nm