CHAPTER 6 Operation
6-40 Sysmex XE-5000 Instructions for Use
Revised July 2007
4. After the READY LED turns off and buzzer stops, remove
the CELLCLEAN.
5. The shutdown sequence in the Main Unit will start.
6. When the shutdown sequence is completed, the message
“Turn OFF Main Unit” will appear.
7. If you wish to complete the analysis, turn OFF the Main
Unit power in the current status.
Warning!
CELLCLEAN is a strong alkaline detergent. Pay
careful attention not to let it adhere to your body or
clothes. If CELLCLEAN adheres to your body or
clothes, wash it away immediately with plenty of water.
Otherwise, it can damage your skin or clothes.
Caution!
When removing the CELLCLEAN container, pull it
down straight not to bend the manual aspiration
pipette.
Otherwise CELLCLEAN may scatter.
Note:
The manual aspiration pipette is automatically rinsed,
and does not need to be wiped manually.
Note:
If you wish to continue an analysis with the Main Unit
left ON, select “Restart” on the function menu of the
Shutdown Completion screen.
After executing an automatic rinse and background
check, the system returns to the READY status.