FUNCTION SETTING USING SETTINGS SCREEN
SETTING THE THRESHOLD K VALUE
The threshold K value of the IOL Selection map is set.
In the following cases, the character color is changed based on the set threshold K value:
When the K value is within the set value range
When the K value is outside the set value range
The threshold value can be set between 0.00 and 100.00 in 0.01 unit.
Before shipment, the default setting is "0.50."
1 On the SETTINGS screen, tap the of the set item display "K Read-
ings" of "DISPLAY."
2 Enter the value by the ten-key display of the .
3 Setting is complete by pressing the .
SETTING THE THRESHOLD VALUE OF CORNEAL SPHERICAL ABERRATION
The threshold value of corneal spherical aberration in the IOL Selection map is set:
In the following case, the character color is changed based on the set threshold value of
corneal spherical aberration:
When the measurement value is below the set value
The threshold value can be set between 0.00 and 100.00 in 0.01 unit.
Before shipment, the default setting is "0.100."
1 On the SETTINGS screen, tap the of the set item display "Corneal
SA" of "DISPLAY."
2 Enter the value by the ten-key display of the .
3 Setting is complete by pressing the .
SETTING THE THRESHOLD VALUE OF CORNEAL ASTIGMATISM
The Threshold Value of Corneal Astigmatism in the IOL Selection map is set:
In the following cases, the character color is changed based on the set threshold value of
corneal astigmatism:
When the measurement value is within the set value range ( range)
When the measurement value is outside the set value range ( range)
The threshold value can be set between 0.00 and 100.00 in 0.01 unit.
Before shipment, the default setting is "1.5."
1 On the SETTINGS screen, tap the of the set item display "Corneal
Astig." of "DISPLAY."
2 Enter the value by the ten-key display of the .
3 Setting is complete by pressing the .
Set Condition button
Setting Change button
OK button
Set Condition button
Setting Change butto
n
OK button
Set Condition button
Setting Change butto
n
OK button