157
FUNCTION SETTING USING SETTINGS
SETTING RMS THRESHOLD 1 (4mm) (THRESHOLD 1 < THRESHOLD 2)
The threshold 1 of RMS (4mm) in the IOL selection map is set.
Based on the threshold 1 of RMS (4mm) and the threshold 2 of RMS (4mm), the color is dif-
ferent according to the followings.
The measurement value is the threshold 1 of RMS (4mm) or lower
The measurement value exceeds the threshold 1 of RMS (4mm) and the RMS threshold 2
or lower.
The measurement value is the threshold 2 of RMS (4mm) or higher
Threshold 1 of RMS (4mm): the lowest threshold set for RMS
Threshold 2 of RMS (4mm): the highest threshold set for RMS
Setting is available with the increment of 0.01 from 0.00 to 100.00.
The default setting is 0.30.
1 Tap the in the "Threshold 1 of RMS (4mm)" in the setting item
"Display Setting".
2 With the numeric keypad of the , input numbers.
3 Press to complete the setting.
SETTING RMS THRESHOLD 2 (4mm) (THRESHOLD 1 < THRESHOLD 2)
Threshold 2 of RMS (4mm) in the IOL Selection map is set.
Setting is available with the increment of 0.01 from 0.00 to 100.00.
The default setting is 0.60.
1 Tap the in the "Threshold 2 of RMS (4mm)" in the setting item
"Display Setting".
2 With the numeric keypad of the , input numbers.
3 Press to complete the setting.
SETTING RMS THRESHOLD 1 (6mm) (THRESHOLD 1 < THRESHOLD 2)
Threshold 1 of RMS (6mm) in the IOL Selection map is set.
Based on the threshold 1 of RMS (6mm) and the threshold 2 of RMS (6mm), the color is dif-
ferent according to the followings.
The measurement value is the threshold 2 of RMS (6mm) or lower
The measurement value exceeds the threshold 1 of RMS (6mm) and the RMS threshold 2
or lower.
The measurement value is the threshold 2 of RMS (6mm) or higher
Threshold 1 of RMS (6mm): the lowest threshold set for RMS
Threshold 2 of RMS (6mm): the highest threshold set for RMS
Setting is available with the increment of 0.01 from 0.00 to 100.00.
The default setting is 0.50.
1 Tap the in the "Threshold 1 of RMS (6mm)" in the setting item
"Display Setting".
2 With the numeric keypad of the , input numbers.
3 Press to complete the setting.
Set Condition button
Setting Change button
OK button
Set Condition button
Setting Change butto
n
OK button
Set Condition button
Setting Change butto
n
OK button