R3132 Series Spectrum Analyzer Operation Manual
2.3.6 Pass/Fail Judgments Using the Limit line Function
2-79
2.3.6 Pass/Fail Judgments Using the Limit line Function
Pass/fail judgments for traces on the screen can easily be made by storing the upper (Line 1) and lower
(Line 2) limit values using the limit line function.
Power on
1. Turn the power on.
Connecting the input signal cable
2. Connect the calibration signal used in the measurement.
Initialization
This resets the current settings to the factory defaults or user-defined presets.
3. Press SHIFT and CONFIG(PRESET).
This sets the analyzer to its presets values.
Setting the measurement conditions
This changes the analyzer settings so that the input signal is displayed more clearly.
4. Press FREQ, 3, 0 and MHz.
The center frequency is set to 30 MHz.
5. Press SPAN, 2, 0 and MHz.
A frequency span of 20 MHz is set.
6. Press LEVEL, 0 and GHz(+dBm).
The reference level is set to 0 dBm.