R3132 Series Spectrum Analyzer Operation Manual
3.3.7 DISPLAY Key (Line and Window)
3-31
3.3.7 DISPLAY Key (Line and Window)
This section describes the menu displayed when the
DISPLAY
key is pressed.
Disp Line ON/OFF Toggles the display line indication on or off. The display line is
used as a base line when comparing trace levels.
ON: Turns the display line on. The display line position
can be adjusted under this setting.
OFF: Turns the display line off.
Ref Line ON/OFF Toggles the reference line indication on or off. The reference line
is used as a base line to which the level value is relative.
ON: Displays the reference line. The reference line posi-
tion can be adjusted under this setting.
OFF: Turns the reference line off.
Measuring Window Displays the Window menu.
Window ON/OFF Turns the measuring window display on or off.
ON: Displays the measuring window.
OFF: Erases the measuring window.
Window Position Allows you to set the measuring window position.
Window Width Allows you to set the measuring window width.
Window Sweep ON/OFF Turns the window sweep function on or off.
ON: Sweeps the range specified by the measuring window.
OFF: Sweeps the range specified by the span.
Marker Couple ON/OFF Turns the marker couple function on or off.
ON: Marker search is limited to the inside of the measuring
window.
OFF: Turns the marker couple function off. The whole screen
is searched.
Multi Screen Displays the Multi Scrn menu.