R3132 Series Spectrum Analyzer Operation Manual
5.2.8 Input Attenuator Switching Accuracy (R3132/32N/62)
5-31
Table 5-11 Input Attenuator Switching Accuracy Test Setting for R3132/32N
Table 5-12 Input Attenuator Switching Accuracy Test Setting for R3162
R3132/32N
External Attenuator
Input Attenuator Reference Level
0 dB -40 dBm 50 dB
10 dB -30 dBm 40 dB
20 dB -20 dBm 30 dB
30 dB -10 dBm 20 dB
40 dB 0 dBm 10 dB
50 dB +10 dBm 0 dB
R3162
External Attenuator
Input Attenuator Reference Level
0 dB -40 dBm 50 dB
5 dB -35 dBm 45 dB
10 dB -30 dBm 40 dB
15 dB -25 dBm 35 dB
20 dB -20 dBm 30 dB
25 dB -15 dBm 25 dB
30 dB -10 dBm 20 dB
35 dB -5 dBm 15 dB
40 dB 0 dBm 10 dB
45 dB +5 dBm 5 dB
50 dB +10 dBm 0 dB