R3132 Series Spectrum Analyzer Operation Manual
1.1 Product Description
1-1
1 INTRODUCTION
This chapter provides the following information:
• Product description
• A list of standard accessories and power cable options
• Options
• Operating environment
• How to verify that the spectrum analyzer is functioning properly
• How to clean, store, and transport the spectrum analyzer
• Calibration
• Replacing Parts with Limited Life
1.1 Product Description
The R3132 Series spectrum analyzer uses the synthesized local method and allow the user highly stable and
highly precise spectrum analysis.
The key features of the R3132 Series spectrum analyzer are listed below.
(1) Frequency Range: 9 kHz to 3 GHz (R3132/N)
9 kHz to 8 GHz (R3162)
9 kHz to 26.5 GHz (R3172)
9 kHz to 40 GHz (R3182)
Frequency span: Zero, 1 kHz to 3 GHz (R3132/N)
Zero, 1 kHz to 8 GHz (R3162)
Zero, 1 kHz to 26.5 GHz (R3172)
Zero, 1 kHz to 40 GHz (R3182)
(2) High-speed and high-precision sweep
Frequency span accuracy: 1% or less
Sweep time: 20 ms
(3) Wide dynamic range
Maximum input level: +30 dBm
Built-in pre-amplifier: -132 dBm noise level
(4) Frequency counter function with a resolution of 1Hz.
(5) A power measurement function useful for evaluating radio instruments using measurements such as
occupied bandwidth (OBW), adjacent channel power (ACP), channel power, etc.
(6) An auto tuning function that searches for a signal with the maximum input level.
(7) Save and recall functions which you can use to store measurement conditions and data in TEXT for-
mat.
(8) A 3.5-inch floppy disk drive which you can use to save screen images in BMP format.
(9) Support for ESC/P, ESC/P-R and PCL compatible printers.