R3132 Series Spectrum Analyzer Operation Manual
2.3.7 Measurements Using TG (OPT74)
2-85
2.3.7 Measurements Using TG (OPT74)
Band-pass filter characteristics with a passband of approximately 270 MHz, are measured (both the inser-
tion loss and bandwidth are measured).
CAUTION: UNCAL messages, displayed when measuring frequency characteristics using this function, do not
affect measurement results.
Setup
1. Connect the unit under test as shown in Figure 2-85.
Figure 2-85 Setup for TG Measurements
Power on
2. Turn the power on.
Initialization
This resets the current settings to the factory defaults.
3. Press SHIFT and CONFIG(PRESET).
The default settings have now been reset.
Setting the measurement conditions
This changes the analyzer settings so that the input signal is displayed more clearly.
4. Press FREQ, 2, 7, 0 and MHz.
A center frequency of 270 MHz is set.
5. Press SPAN, 1, 0, 0 and MHz.
A frequency span of 100 MHz is set.
6. Press LEVEL, 0 and GHz(+dBm).
The reference level is set to 0 dBm.
Through
R3132 Series Spectrum analyzer